Shopping Cart

No products in the cart.

AAMI 14117 2019

$162.84

ANSI/AAMI/ ISO 14117:2019 – Active implantable medical devices-Electromagnetic compatibility-EMC test protocols for implantable cardiac pacemakers, implantable cardioverter defibrillators and cardiac resynchronization devices

Published By Publication Date Number of Pages
AAMI 2019 152
Guaranteed Safe Checkout
Category:

If you have any questions, feel free to reach out to our online customer service team by clicking on the bottom right corner. Weโ€™re here to assist you 24/7.
Email:[email protected]

Specifies a comprehensive test methodology for the evaluation of the electromagnetic (EM) compatibility of active implantable cardiovascular devices. The devices addressed by this standard include those that provide one or more therapies for bradycardia, tachycardia, and cardiac resynchronization. This document details test methods appropriate for the interference frequencies at issue. It specifies performance limits or requires disclosure of performance in the presence of EM emitters, where indicated.

PDF Catalog

PDF Pages PDF Title
1 ANSI/AAMI/ISO 14117:2019; Active implantable medical devicesโ€”Electromagnetic compatibilityโ€”EMC test protocols for implantable cardiac pacemakers, implantable cardioverter defibrillators and cardiac resynchronization devices
3 Title page
4 AAMI Standard
Copyright information
5 Contents
6 Committee representation
7 Background of ANSI/AAMI adoption of ISO 14117:2019
8 Foreword
10 Introduction
13 1 Scope
2 Normative references
3 Terms, definitions, symbols and abbreviated terms
3.1 Terms and definitions
15 3.2 Acronyms and abbreviations
Table 1 โ€” List of acronyms and abbreviations
17 4 Test requirements for the frequency band 0 Hz โ‰ค ฦ’ โ‰ค 3 000 MHz
4.1 General requirements for all devices
4.2 Induced lead current
4.2.1 General requirements
18 4.2.2 Pacemakers and CRT-P devices
Figure 1 โ€” Test signal 2
19 Figure 2 โ€” Test setup for measurement of induced current
20 Figure 3 โ€” Connection to a single-channel unipolar device
Figure 4 โ€” Connection to a multichannel unipolar device
Figure 5 โ€” Common mode connection to single-channel bipolar device
21 Figure 6 โ€” Differential mode connection to single-channel bipolar device
Figure 7 โ€” Common mode connection to multichannel bipolar device
Figure 8 โ€” Differential mode connection to multichannel bipolar device
22 Table 2 โ€” Spurious injection current limits
4.2.3 ICDs and CRT-D devices
4.2.3.1 Test requirements
4.2.3.2 Measurement of current injected through sense/pace terminals
23 Figure 9 โ€” Common mode connection to multichannel bipolar device
Figure 10 โ€” Differential mode connection to multichannel bipolar device
4.2.3.3 Measurement of current injected through cardioversion/defibrillation terminals
24 Figure 11 โ€” Common mode connection for cardioversion/defibrillation terminals
Figure 12 โ€” Differential mode connection for cardioversion/defibrillation terminals
25 Table 3a โ€” Spurious injection current limits for sense/pace terminals
Table 3b โ€” Spurious injection current limits for cardioversion/defibrillation terminals
4.3 Protection from persisting malfunction attributable to ambient electromagnetic fields
4.3.1 General requirements
4.3.2 Pacemaker and CRT-P devices
4.3.2.1 Malfunction due to electromagnetic interference in the frequency range of 16,6 Hz to 10 MHz
26 Table 4 โ€” Peak-to-peak amplitudes Vpp in the range 16,6 Hz to 10 MHz
Figure 13 โ€” Test setup to check for induced malfunction
27 Figure 14 โ€” Connection to a single-channel unipolar device
Figure 15 โ€” Connection to a multichannel unipolar device
Figure 16 โ€” Common mode connection to a single-channel bipolar device
Figure 17 โ€” Differential mode connection to a single-channel bipolar device
28 Figure 18 โ€” Common mode connection to a multichannel bipolar device
Figure 19 โ€” Differential mode connection to a multichannel bipolar device
4.3.2.2 Malfunction due to electromagnetic interference in the frequency range of 10 MHz to 385 MHz
Figure 20 โ€” Test signal for frequencies between 10 MHz and 385 MHz
29 Figure 21 โ€” Test setup to check for malfunction at high frequency
30 Figure 22 โ€” Connection to a unipolar device
Figure 23 โ€” Connection to a bipolar device
4.3.2.3 Malfunction due to electromagnetic interference in the frequency range of 385 MHz to 3 000 MHz
4.3.3 ICDs and CRT-D devices
4.3.3.1 Malfunction due to electromagnetic interference in the frequency range of 16,6 Hz to 10 MHz
4.3.3.1.1 Test equipment and signal
31 Table 5 โ€” Peak-to-peak amplitudes Vpp in the range 16,6 Hz to 10 MHz
4.3.3.1.2 Malfunction because of electrical interference on the sense or pace terminals
32 Figure 24 โ€” Common mode connection for multichannel bipolar devices
Figure 25 โ€” Differential mode connection for multichannel bipolar devices
4.3.3.1.3 Malfunction because of electromagnetic interference on the cardioversion/ defibrillation terminals
Figure 26 โ€” Test setup to check for induced malfunction attributable to voltages induced on cardioversion/defibrillation terminals
33 Figure 27 โ€” Common mode connection for cardioversion/defibrillation terminals
Figure 28 โ€” Differential mode connection for cardioversion/defibrillation terminals
4.3.3.2 Malfunction due to electromagnetic interference in the frequency range of 10 MHz to 385 MHz
4.3.3.2.1 Test equipment and signal
34 Figure 29 โ€” Test signal for frequencies between 10 MHz and 385 MHz
4.3.3.2.2 Malfunction because of electrical interference on the sense or pace terminals
35 Figure 30 โ€” Test set-up to check for induced malfunction at high frequency
Figure 31 โ€” Connection of the DUT
36 4.3.3.2.3 Malfunction because of electromagnetic interference on the cardioversion/ defibrillation terminals
4.3.3.3 Malfunction due to electromagnetic interference in the frequency range of 385 MHz to 3 000 MHz
4.4 Protection from malfunction caused by temporary exposure to CW sources
4.4.1 Pacemaker and CRT-P device response to temporary continuous wave sources in the frequency range 16,6 Hz to 167 kHz
37 Figure 32 โ€” Test setup to characterize DUT performance while subject to interference
38 4.4.2 ICDs and CRT-D devices
39 4.5 Protection from sensing EMI as cardiac signals
4.5.1 General requirements
40 4.5.2 Protection from sensing EMI as cardiac signals in the frequency range of 16,6 Hz to 150 kHz
4.5.2.1 Pacemakers and CRT-P devices
Table 6 โ€” Peak-to-peak amplitudes Vpp in the range of 16,6 Hz to 150 kHz
41 Figure 33 โ€” Test signal 2: Used in the range of 1 kHz to 150 kHz
4.5.2.2 ICDs and CRT-D devices
42 Table 7โ€” Peak-to-peak amplitudes Vpp in the range of 16,6 Hz to 150 kHz
Figure 34 โ€” Test signal 2: Used in the range of 1 kHz to 150 kHz
44 4.5.3 Protection from sensing EMI as cardiac signals in the frequency range of 150 kHz to 10 MHz
4.5.3.1 Pacemakers and CRT-P devices
Figure 35 โ€” Test signal for frequencies 150 kHz to 10 MHz
Table 8 โ€” Peak-to-peak test signal amplitudes Vpp in the range of 150 kHz to 10 MHz
45 4.5.3.2 ICDs and CRT-D devices
Table 9 โ€” Peak-to-peak test signal amplitudes Vpp in the range of 150 kHz to 10 MHz
Figure 36 โ€” Test signal for frequencies 150 kHz to 10 MHz
47 4.5.4 Protection from sensing EMI as cardiac signals in the frequency range of 10 MHz to 385 MHz
4.5.4.1 Pacemakers and CRT-P devices
48 Figure 37 โ€” Test setup to check for malfunction at high frequency
49 Figure 38 โ€” Connection to a unipolar device
4.5.4.2 ICDs and CRT-D devices
4.6 Protection from static magnetic fields of flux density up to 1 mT
4.6.1 General requirements
4.6.2 Pacemakers and CRT-P devices
50 Figure 40 โ€” Test setup for magnetostatic measurements
4.6.3 ICDs and CRT-D devices
4.7 Protection from static magnetic fields of flux density up to 50 mT
4.7.1 General requirements
51 4.7.2 Pacemakers and CRT-P devices
4.7.3 ICDs and CRT-D devices
4.8 Protection from AC magnetic field exposure in the range of 1 kHz to 140 kHz
4.8.1 General requirements
4.8.2 Pacemakers and CRT-P devices
Table 10 โ€” Sinusoidally modulated magnetic field strengths
52 Figure 41 โ€” Loop configuration for varying magnetic field test
4.8.3 ICDs and CRT-D devices
4.9 Test requirements for the frequency range of 385 MHz โ‰ค ฦ’ โ‰ค 3 000 MHz
4.9.1 General requirements
53 4.9.2 Test setup
4.9.2.1 Test environment
4.9.2.2 Torso simulator in Annex G
Table 11 โ€” Requirements for the test setup
4.9.2.3 Device under test and lead positioning in torso simulator
4.9.2.4 Interference signal generation
54 4.9.2.5 Parameter programming
4.9.2.6 Monitoring of device activity
4.9.2.7 Simulated cardiac signal injection
4.9.3 Test procedure
4.9.3.1 Protection from proximity fields at 15 cm separation distance
56 4.9.3.2 Optional characterization testing
4.9.4 Performance criteria
4.9.4.1 Single-chamber pacing modes
4.9.4.2 Multi-chamber pacing modes
57 4.9.4.3 Antitachyarrhythmia modes
4.10 Transient exposure to stationary low-frequency electromagnetic field sources in the frequency range 16,6 Hz to 167 kHz
5 Testing above frequency of 3 000 MHz
58 6 Protection of devices from EM fields encountered in a therapeutic environment
6.1 Protection of the device from damage caused by high-frequency surgical exposure
6.1.1 General requirements
6.1.2 Pacemakers and CRT-P devices
Figure 42 โ€” Test setup for protection of the device from high-frequency currents caused by high-frequency surgical equipment
59 Table 12 โ€” Test signal characteristics
6.1.3 ICDs and CRT-D devices
6.2 Protection of the device from damage caused by external defibrillators
6.2.1 General requirements
6.2.2 Pacemakers and CRT-P devices
60 Figure 43 โ€” Damped sinus waveform
Figure 44 โ€” Circuit for generating a damped sinus defibrillation waveform for Test 1
61 Figure 45 โ€” Timing sequence used in Tests 1 and 2
62 Figure 46 โ€” Test setup for Test 2 (using a truncated exponential defibrillation waveform)
Figure 47 โ€” Resistor network for Tests 1 and 2
Table 13 โ€” Resistor network parameters
63 Figure 48 โ€” Biphasic defibrillation waveform for Test 2
6.2.3 ICDs and CRT-D devices
7 Additional accompanying documentation
7.1 Disclosure of permanently programmable sensitivity settings
7.2 Descriptions of reversion modes
7.3 Known potential hazardous behaviour
7.4 Minimum separation distance from hand-held transmitters
64 Annex A (informative) Rationale
A.1 Rationale for test requirements for the frequency band 0 Hz โ‰ค ฦ’ < 385 MHz (see 4.1 to 4.8)
69 Figure A.1 โ€” Maximum allowed current when injected to the heart
72 A.2 Rationale for test requirements for the frequency band 385 MHz โ‰ค ฦ’ โ‰ค 3 000 MHz (see 4.9)
A.2.1 Rationale for DUT reference point
A.2.2 Rationale for the RF modulation
A.2.3 Rationale for the optional characterization testing
A.2.4 Rationale for test power levels
73 Figure A.2 โ€” Dipole net power measurements (dipole spacing = 2,5 cm) conducted for ANSI/AAMI PC69:2000
74 A.2.5 Rationale for lead configuration
A.2.6 Rationale for device programmed parameters
75 A.3 Rationale for sample size
A.4 Rationale for test requirements in Clause 6
A.4.1 Protection of the device from damage caused by high-frequency surgical exposure
A.4.2 Protection of the device from damage caused by external defibrillators
76 A.4.3 Test signal modulation format
77 Annex B (informative) Rationale for test frequency ranges
78 Annex C (informative) Code for describing modes of implantable generators
C.1 The code
Table C.1 โ€” NASPE/BPEG generic (NBG) pacemaker code
79 Table C.2 โ€” NASPE/BPEG defibrillator (NBD) code
80 Annex D (normative) Interface circuits
Figure D.1 โ€” Tissue-equivalent interface circuit for current measurements
Table D.1 a) Component values for Figure D.1
81 Table D.1 b) Component values for Figure D.1
Figure D.2 โ€” Tissue-equivalent interface circuit to check for malfunction
82 Table D.2 โ€” Component values for Figure D.2
Figure D.3 โ€” Tissue-equivalent interface circuit to check for malfunction caused by voltages induced on cardioversion/defibrillation terminals
83 Table D.3 โ€” Component values for Figure D.3
Figure D.4 โ€” Low-pass filter used to attenuate the 500 kHz component of a test signal (see 4.2.2, 4.2.3, and Annex E)
Table D.4 โ€” Component values for Figure D.4
84 Figure D.5 โ€” Injection network
Table D.5 โ€” Component values for Figure D.5
86 Annex E (informative) Selection of capacitor Cx
87 Figure E.1 โ€” Example amplitude at point D and C of the tissue-equivalent interface (CX selected for 5 000 Hz corner frequency)
88 Figure E.2 โ€” Test to check for spurious low-frequency noise and to determine the value of Cx
89 Annex F (normative) Calibration of the injection network (Figure D.5)
90 Table F.1 โ€” Calibration signal amplitude
91 Annex G (normative) Torso simulator
G.1 Torso simulator
G.2 Top grid
G.3 Cutout
G.4 Bottom grid
G.5 Torso simulator electrodes
92 G.6 Illustrations
93 Figure G.1 โ€” Torso simulator
94 Figure G.2 โ€” Test setup
95 Annex H (normative) Dipole antennas
H.1 Resonant dipole
Table H.1 โ€” Dipole description
96 Figure H.1 โ€” Example of a dipole antenna
97 Annex I (normative) Pacemaker/ICD programming settings
I.1 General
I.2 Pacemaker
I.2.1 Parameters
Table I.1 โ€” Pacemaker parameters
I.2.2 Diagnostic settings
98 I.3 ICD
I.3.1 Parameters
Table I.2 โ€” Tachycardia device parameters
I.3.2 Diagnostic settings
99 I.4 Other operating modes or parameters not implied in this document
100 Annex J (normative) Simulated cardiac signal
J.1 Heart simulated signal
Figure J.1 โ€” Simulated cardiac signal
101 Annex K (normative) Calculation of net power into dipole antenna
K.1 Calculation of net dipole power
K.1.1 Calculation of forward dipole power (dBm)
K.1.2 Conversion of forward dipole power from dBm to milliwatts
K.1.3 Calculation of reflected dipole power (dBm)
102 K.1.4 Conversion of reflected dipole power from dBm to milliwatts
K.1.5 Calculation of net dipole power (mW)
K.2 Measurement of factors for net power calculations
K.2.1 DCF โ€” Directional coupler forward port coupling factor
K.2.2 DCR โ€” Directional coupler reflected port coupling factor
103 K.2.3 ACA antenna cable attenuation
106 Annex L (informative) Loop area calculations
L.1 Purpose
L.2 Procedure
107 Figure L.1 โ€” Simulated right pectoral dual-chamber pacemaker X-ray
108 Figure L.2 โ€” Simulated left pectoral dual-chamber ICD X-ray
L.3 Results
109 Table L.1 โ€” Pacemaker systems
110 Table L.2 โ€” ICD systems
111 L.4 Summary โ€” Geometrical lead loop area
L.5 Effective induction area
112 Figure L.3 โ€” Effective induction area of an open wire loop inside a conductive medium
L.6 Consideration of Lead Loop Area based on CRT-P/CRT-D devices and Left Ventricular lead placement
113 Annex M (informative) Correlation between levels of test voltages used in this document and strengths of radiated fields
117 Figure M.1 โ€” Comparison of induced voltages resulting from ICNIRP and IEEE H-field general public reference levels and the test limits of this document
118 Figure M.2 โ€” Comparison of magnetic field general public reference exposure values from ICNIRP and IEEE with equivalent magnetic field immunity from ISO 14117
119 Figure M.3 โ€” Induced voltage zones
120 Figure M.4 โ€” Magnetic field zones
121 Annex N (informative) Connections to DUTs having ports with more than two electrode connections
N.1 Introduction
N.2 New nomenclature of ports and electrodes
122 Figure N.1 โ€” Schematic representation of ports and electrodes for a generic multi-port multi-polar device; the order of the ports, and electrodes in each port shown in the picture is arbitrary
123 N.3 Examples of application of the nomenclature to typical DUTs
Figure N.2 โ€” Example of DR pacemaker device according to the new nomenclature
125 Figure N.3 โ€” Example of DF-1 CRT-D device according to the new nomenclature
127 Figure N.4 โ€” Example of DF-4 IS-4 CRT-D device according to the new nomenclature
N.4 Generalization of tissue-equivalent circuit and test setup
N.4.1 Nomenclature
128 Figure N.5 a) Tissue-equivalent interface extension to multi-port multi-polar devices. Fp,e indicates a network output paired to the low-voltage electrode โ€˜eโ€™ of device port โ€˜pโ€™
130 N.4.2 Unipolar Injection Configuration โ€” Tests 4.3/4.4/4.5 (up to 10 MHz)
Figure N.6 โ€” Extension of Unipolar Injection Configuration for multi-port multi-polar devices- applicable to tests 4.3/4.4/4.5 (up to 10 MHz)
131 Figure N.7 โ€” Extension of Unipolar Injection Configuration for multi-port multi-polar devices. Example of a DR pacemaker
Figure N.8 โ€” Extension of Unipolar Injection Configuration for multi-port multi-polar devices. Example of a DF-1 CRT-D
132 Figure N.9 โ€” Extension of Unipolar Injection Configuration for multi-port multi-polar devices. Example of a DF-4 IS-4 CRT-D
133 N.5 Multipolar Common Mode Injection Configuration โ€” Tests 4.3/4.4/4.5 (up to 10 MHz)
Figure N.10 โ€” Extension of Multipolar Common Mode Injection Configuration for multi-port multi-polar devices- applicable to tests 4.3/4.4/4.5 (up to 10 MHz)
134 Figure N.11 โ€” Extension of Multipolar Common Mode Injection Configuration for multi-port multi-polar devices. Example of a DR pacemaker
135 Figure N.12 โ€” Extension of Multipolar Common Mode Injection Configuration for multi-port multi-polar devices. Example of a DF-1 CRT-D
136 Figure N.13 โ€” Extension of Multipolar Common Mode Injection Configuration for multi-port multi-polar devices. Example of a DF-4 IS-4 CRT-D
137 N.6 Multipolar Differential Mode Injection Configuration โ€” Tests 4.3/4.4/4.5 (up to 10 MHz)
Figure N.14 โ€” Extension of Bipolar Differential Mode Injection Configuration for multi-port multi-polar devices- applicable to tests 4.3/4.4/4.5 (up to 10 MHz)
138 Figure N.15 โ€” Extension of Bipolar Differential Mode Injection Configuration for multi-port multi-polar devices. Example of a DR pacemaker
140 Figure N.16 โ€” Extension of Bipolar Differential Mode Injection Configuration for multi-port multi-polar devices. Example of a DF-1 CRT-D
143 Figure N.17 โ€” Extension of Bipolar Differential Mode Injection Configuration for multi-port multi-polar devices. Example of a DF-4 IS-4 CRT-D
145 Annex O (informative) Example method for evaluation of transient and permanent malfunction of a CIED due to temporary exposure to low frequency (<167 kHz) electromagnetic fields
O.1 General considerations
O.2 Pacemakers and CRT-P devices
146 Figure O.1 โ€” Test signal 1: CW exposure
147 Figure O.2 โ€” Test signal 2: Burst-modulated exposure (for illustration purposes modulation parameters B and T are not scaled)
148 O.3 ICDs and CRT-D devices
O.4 Rationale for test signal modulation
149 Figure O.3 โ€” Biot-Savart law for a circular current loop emitter; Bz is the B-field component along the z-axis perpendicular to the current loop
150 Bibliography
AAMI 14117 2019
$162.84